Agent consumption with the Zeus® in the automated closed circuit anesthesia mode with O2/air mixtures

Abstract

Background

Earlier software versions of the Zeus® (Lübeck, Dräger, Germany) failed to provide true closed circuit anesthesia (CCA) conditions. We examined whether the latest software (SW 4.03 MK 04672–00) achieves this goal.

Methods

In 8 ASA I–III patients, the CCA mode of the Zeus® was used to maintain the inspired O2 (FIO2) and end-expired sevoflurane % (FAsevo) at 50 and 1.8%, respectively. The fresh gas flow (FGF) of O2 and air and the sevoflurane injection rate (=Vinjsevo, mL liquid sevo/h) were videotaped from the control screen and entered offline into a spreadsheet. Cumulative sevoflurane usage during early wash-in (=0-1 min, CDsevo0-1), late wash-in (=1-5 min, CDsevo1-5), and maintenance (=5-60 min, CDsevo5-60) was calculated, and Vinjsevo between 1 and 60 min was compared with published uptake data.

Results

FAsevo reached 1.8% within 101 (23) sec. CDsevo0-1 was between 1.24 (0.03) and 3.01(0.25) mL (a range is provided because no absolute Vinjsevo values were displayed once Vinjsevo was > 100 mL/h, which occurred between 15 ± 2 and 46 ± 6 sec). CDsevo1-5 was 0.81 (0.37) mL, and CDsevo5-60 was 4.63 (0.94) mL. The Vinjsevo pattern between 1 and 60 min matched previously published uptake data. Brief high FGF periods were used to maintain the target FIO2, and to refill the reservoir bag after external pressure had been applied to the abdomen; subsequent “spikes” wasted 0.08-0.19 mL and 0.14-0.49 mL sevoflurane (1-3% and 3-9% of total agent usage between 1 and 60 min, respectively).

Conclusion

Under the conditions specified, the Zeus® approaches CCA conditions so closely that further reductions in agent usage would have minimal economic significance.

Background

Closed-circuit anesthesia (CCA) conditions exist when the amount of agent and carrier gas administered match the amounts needed to prime the circle breathing system, taken up by the patient, and lost via leaks. The initial version of the only commercially available automated closed circuit anesthesia (CCA) machine at the time of the study, the Zeus® (Lübeck, Dräger, Germany), failed to reduce agent usage to levels approaching CCA conditions, because FGF usage during the first minutes had been programmed excessively high [1, 2]. New software now limits the duration and the pattern of the initial high FGF period and has limited the number of intermittent high FGF flushes during maintenance (intended to reduce unwanted gases like N2, compound A, carbon monoxide, methane). We hypothesize this should improve the performance by reducing agent and carrier gas usage to near closed-circuit conditions, and therefore studied the sevoflurane usage with the latest software version, SW 4.03 MK 04672–00, in the automated CCA mode.

Methods

After obtaining IRB approval (IRB of the Kliniek Sint Jan, Brussels, Belgium; Human studies number OM072 Ref 2012.115) and written informed consent, 8 ASA I –III patients presenting for abdominal or breast surgery were enrolled. The patients’ age, height, and weight were recorded. One hour prior to surgery, 0.5 mg alprazolam p.o. was administered. After preoxygenation with 8 L/min O2 by facemask, anesthesia was induced with sufentanil (0.1 mg/kg) and propofol (3 mg/kg). Intubation of the trachea was facilitated by rocuronium (0.5 mg/kg) or cisatracurium (0.1-0.15 mg/kg).

After connecting the endotracheal tube to the anesthesia circuit, ventilation was mechanically controlled with the Zeus® anesthesia machine (software version SW 4.03 MK 04672–00), with tidal volume = 500 mL, respiratory rate = 10/min, and I:E ratio = 1:2. The attending anesthesiologist was allowed to adjust ventilation to maintain normocapnia. Anesthesia was maintained with sevoflurane in O2/air using the automated CCA mode. Time zero was the time at which the target FIO2 was set at 50% and the target end-expired sevoflurane concentration (FAsevo) at 1.8%. From this point on, the machine automatically adjusted the O2 and air FGF as well as the sevoflurane liquid injection rate (Vinjsevo) to attain and maintain the specified targets. Additional sufentanil and rocuronium or cisatracurium administration were left at the discretion of the attending anesthesiologist. The study arbitrarily lasted 60 min.

Because we could not obtain software to download the data with a resolution sufficient for the purposes of this study (at least every second), the monitor screen was video-recorded. The values of the following parameters were entered offline every second into a spreadsheet (amounting to 3600 entries per parameter per patient): (1) time; (2) the O2 and air FGF and Vinjsevo; and (3) the resulting inspired and end-expired O2, CO2, and sevoflurane concentrations (FIO2, FAO2, FICO2, FACO2, FIsevo, and FAsevo, respectively).

Cumulative sevoflurane consumption was obtained by integrating the area under the curve of Vinjsevo. The Vinjsevo value is displayed on the Zeus screen. This value is the average of multiple, very precisely dosed liquid “pulses” injected per second from the so-called DIVA cassette, basically a fuel-injector from a car engine that injects liquid agent with precise dosing volume of 3 to 50 microL (Wilfried Buschke, Dräger, personal communication).

To assess how closed the Zeus® worked, we compared the Vinjsevo pattern with previously published sevoflurane uptake data derived from several different sources: closed circuit liquid injection [3, 4], indirect calorimetry (gas balances within the circuit) [5, 6], and the reverse Fick method [5, 6]. A proportional correction was applied to account for differences in FAsevo. Patient demographics, cumulative agent usage, and times are presented as mean ± standard deviation. Because other data (FGF, Vinjsevo, FAsevo) are not normally distributed, they are presented as median and quartiles.

Results

Patient age, height, and weight were 58 ± 13 years, 164 ± 9 cm, and 77 ± 15 kg. Median total FGF (Figure 1a) remained between 150 and 200 mL/min throughout the procedure, with the composition changing progressively from air to O2 (Figure 1b,c) as the FIO2 decreased slowly towards the target 50% (Figure 1d). High FGF bursts were few (Table 1), and were mainly used to increase FIO2 to its target or to fill the breathing bag after pressure had been exerted on the abdomen or thorax by the surgical team that resulted in loss of gas from the system.

Figure 1
figure 1

Carrier gas characteristics. Total fresh gas flow remained below 200 mL/min (a), but the composition of the fresh gas changed over time: air fresh gas flow (b) decreased, while O2 fresh gas flow (c) increased. The inspired O2 concentration progressively decreased towards the target of 50% (d). Thick lines = median, thin lines = quartiles.

Table 1 Overview of intermittent high fresh gas episodes and ensuing sevoflurane spikes

FAsevo reached 1.8% within 101 ± 23 sec (Figure 2a) and was maintained within 0.1% thereafter (Figure 2b). After the first minute, the Vinjsevo pattern matched previously published uptake data [36] (Figure 2c). Vinjsevo “spikes” followed the high FGF episodes mentioned above, and wasted 0.08-0.19 mL and 0.14-0.49 mL liquid sevoflurane when used to increase FIO2 and after external pressure had been applied to the abdomen, constituting 1-3% and 3-9% of total agent usage between 1 and 60 min, respectively (Table 1).

Figure 2
figure 2

Inspired and end-expired sevoflurane concentrations during wash-in (a) and maintenance (b) (upper and lower line, thick line = mean, thin lines = standard deviation), and corresponding liquid sevoflurane injection rate (mL/h) (c and d, respectively). During the first 5 minutes (c), the liquid sevoflurane injection rate (mL/h) could not be accurately measured between 15 ± 2 and 46 ± 6 sec because Vinjsevo was > 100 mL/h (see text for details). The sevoflurane injection rates match those of previously published sevoflurane uptake data derived from several different sources (d): closed circuit liquid injection (yellow line = Lockwood data, green line = Hendrickx data) [3, 4], indirect calorimetry (gas balances within the circuit = red line) [5, 6], and the reverse Fick method (pink line) [5, 6]. Thick lines = median, thin lines = quartiles.

Cumulative liquid sevoflurane usage was 0.81 ± 0.37 mL between 1–5 min, and 4.63 ± 0.94 mL between 5–60 min. The cumulative sevoflurane usage between 0 and 1 min could not be accurately measured because Vinjsevo was > 100 mL/h between 15 ± 2 and 46 ± 6 sec (Figure 2d); under these circumstances, the Zeus ceases to display values and only displays a message “Vinjsevo > 100 mL/h”. Still, cumulative agent usage can be reported as a range because the maximum Vinjsevo = 300 mL/h, thus the injection rate during these “blackout” episodes has to lie between 100 and 300 mL/h. For example, for a 12 sec blackout period, the cumulative sevoflurane amount ranges between 0.33 to 1 mL liquid. According to these injector limits, cumulative sevoflurane usage between 0 and 1 min was between 1.24 ± 0.03 and 3.01 ± 0.25 mL.

Discussion

While older studies found that with the older algorithms the Zeus® could not reduce agent usage to the amounts needed to prime the system and to replace the amounts taken up by the patient [2], our current data suggest that the newest algorithms applied by the Zeus® have succeeded in reducing agent usage to just 4–12% above CCA conditions [35]. Besides software improvements, other factors may account for this.

First, the algorithms steering the FGF and agent injector during wash-in have been optimized. The Zeus® manages FGF and agent administration to attain and maintain the initial target FA of the agent and carrier gases in such a manner that an acceptable trade-off is made between the speed of reaching the targets (FAO2, FAN2O, and FA agent) and usage of carrier gas and agent. Requests for a higher FIO2 or lower FA agent are considered high priority, necessitating a rapid “step” change and thus a high FGF, which comes at the price of increased agent usage. The response to a request for a lower FIO2 may be allowed more time. Exceptions may include the use of laser in the airway, or the care of neonates where a fast reduction in FIO2 may be required. This can be accomplished by the use of a fast flush button pre-programmed with a FGF according to the user’s preference, but the usage of any concomitantly administered agent will increase. A request for a higher FA agent does not require a high FGF because agent administration and FGF are mechanically uncoupled: agent is injected by the DIVA cassette, and a blower ensures rapid mixing in the circuit. The high FGFs programmed into the early software during initial wash-in have been rewritten: FIO2 is allowed to drift gradually towards the target FIO2 if the initial FIO2 is higher than the target FIO2. Figure 1 illustrates that the average total FGF during the first 5 min varies within a small marge of 150–210 mL/min.

Second, N2O used in previous studies may have increased agent usage compared to N2 (used in this study). If N2O is used, high initial FGF are required to achieve sufficiently high concentrations of N2O in a fairly short time period. Therefore, it is more difficult to maintain CCA from the very beginning of the anesthetic, and agent usage will initially be higher compared to when O2/N2 is used. However, (1) FAN2O may be allowed to rise more gradually, and (2) the effect of N2O on agent consumption is bimodal: after a while, total agent usage doses become lower when N2O is used because a lower agent concentration can be used [7].

Third, few high FGF bursts (flushing) were used during the maintenance phase (Table 1). What purpose do these high FGF bursts serve, to what extent do they increase agent usage, why were they infrequent in our study, and can they be further reduced or eliminated all together? Earlier software versions commanded “routine” intermittent flushing of the breathing system to eliminate unwanted gases like CO, compound A, or methane. However, their clinical relevance has become questionable (with KOH free CO2 absorbents) or entirely irrelevant (with KOH and NaOH free CO2 absorbents). These “routine” bursts have therefore been eliminated, but the concerned anesthesiologist can activate the flush button.

Flushing is also initiated after a certain N2 threshold has been exceeded when N2O/O2 is used, or more precisely after the 10% “balance gas” threshold has been exceeded (N2 continues to be released from slowly equilibrating tissues and thus slowly accumulates in the circuit). The balance gas is calculated as follows: 100% - (FIN2O + FIO2 + FAagent + 7 vol% H2O). Because we used O2/air as the carrier gas, and because the FIO2 was allowed to gradually drift down towards the target FIO2, very few high FGF bursts were needed (n = 5, total wasted carrier gas volume 1.1 ± 0.3 L) (Table 1). But even with the use of O2/air mixtures, the continued release of N2 may cause the FIO2 to eventually drop below its target, initiating a brief O2 flush, which occurred in 4 patients after 45 min (Table 1).

Six FGF bursts were also prompted by a decrease of the pressure in the breathing bag at end-expiration caused by pressure exerted on the abdomen or thorax by the surgical team; these were short-lived (6.3 ± 0.5 sec) and wasted 1.5 ± 0.2 L of circuit gas. The bag is part of the FGF uncoupling system; if the pressure in the bag is not slightly positive at the end of expiration, the system perceives this as lack of sufficient fresh gas, and the FGF is increased. Three other bursts occurred after gas sampling line kinking (for safety reasons no rebreathing is allowed without 2 properly functioning gas analyzers), FAsevo overshoot (a maximum increase of 15% above target is allowed), or post-calibration (even though this did not happen after other calibration periods).

FGF bursts are also used to rapidly attain new targets during the maintenance phase. Demand for a higher FIO2 or lower FA agent in particular are considered high priority requests, necessitating a rapid “step” change and thus a high FGF period, which does come at the price of increased agent usage. The response to a demand for a lower FIO2 is allowed more time (see above). A demand for a higher FA agent does not require a high FGF because agent administration does not depend on FGF (it is injected by the DIVA cassette, and a blower ensures rapid mixing in the circuit). Because we did not change the O2 and FAsevo targets, the number of FGF bursts and thus carrier gas and agent waste were minimized.

We believe our results allow us to conclude that the Zeus decreases agent usage to very-near CCA conditions. During the first minute, the exact amount of liquid sevoflurane usage is unkown, but from the injector limits we can deduct it is between 1.24 ± 0.03 and 3.01 ± 0.25 mL. Future studies might consider weighing the DIVA cassette or determine the amount of agent exhausted towards the scavenging system. Our preliminary data indicate that sevoflurane waste from the exhaust valve under identical study conditions is 0.014 ± 0.004 mL (range 0 – 0.018 mL) liquid sevoflurane during the first 5 min. The first few minutes of agent delivery are crucial to help minimize agent waste. Comparison of usage data of the first few minutes between studies is complicated by differences in circle system configurations and the rate of rise of the agent concentration. During maintenance, the current software version of the Zeus® reduces total FGF and O2 FGF in particular to previously published uptake rates [8]. But the number of these high FGF episodes and their effect on concomitant agent usage has become so small that the excess waste has become negligible vis-à-vis total sevoflurane usage - further reductions would be economically and environmentally insignificant, and technically difficult to achieve.

There are several implications of our findings. First, technology has evolved up to a point where potent inhaled anesthetic agents can be administered with almost no waste. Still, the “purist” can keep agent usage low by minimizing target changes and by accepting that targets be reached gradually [9], principles that apply to any CCA technique. Second, considering that most anesthesiologists still use a maintenance FGF of 1.5 – 2 L/min when using a conventional anesthesia machine [10, 11], automating CCA to maintain FAsevo at 1.8% in O2/air will reduce agent usage during the maintenance phase (0 – 55 min) alone by 367%, from 17.0 mL [12] to 4.63 mL (current study). Yet how this will translate into costs savings is complex, and requires considerable detailed information. For example, one study that compared the costs of the Zeus with the Primus failed to accurately compare the FA agent and the number of FA agent changes, the initial FGFs, and the O2 concentrations, precluding any meaningful conclusions to be made from these results [13]. Another example is the imposed use of high FGF when sevoflurane is used due to concerns for Compound A or CO formation: although the newer CO2 absorbents do not produce these substances, the increased cost of the more expensive and less efficient KOH and NaOH free CO2 absorbent will still be outweighed by the savings made by using less sevoflurane when using an automated CCA machine [14]. Finally, the inhaled anesthetic drugs released during the approximately 200 million anesthetic procedures performed each year globally have a climate impact that is approximately 0.01% of that of the CO2 released from global fossil fuel combustion [15]. Technology now enables us to even reduce this by another order of magnitude, making our contribution to ozone layer destruction and green house effect exceedingly small.

Conclusions

Under the conditions specified in this study, software version SW 4.03 MK 04672–00 has made the automated CCA mode of the Zeus® approach CCA conditions except for brief functional high FGF episodes that result in waste of 3-9% per episode of the total amount of agent administered. Strictly speaking, CCA conditions still were not met 100% of the time, but even under the best of conditions, there always will be some need for intermittent high FGF bursts. It is unlikely this very small amount of waste can be further reduced, and it is likely that any further reduction would be irrelevant. We conclude that the Zeus approaches CCA conditions.

Abbreviations

CCA:

Closed circuit anesthesia

FGF:

Fresh gas flow

FIO2:

Inspired O2 concentration

FAO2:

End-expired O2 concentration

FICO2:

Inspired FICO2 concentration

FACO2:

End-expired CO2 concentration

FIsevo:

Inspired sevoflurane concentration

FAsevo:

End-expired sevoflurane concentration

Vinjsevo:

Sevoflurane injection rate

References

  1. Struys MM, Kalmar AF, De Baerdemaeker LE, Mortier EP, Rolly G, Manigel J, Buschke W: Time course of inhaled anaesthetic drug delivery using a new multifunctional closed-circuit anaesthesia ventilator. In vitro comparison with a classical anaesthesia machine. Br J Anaesth. 2005, 94: 306-317. 10.1093/bja/aei051.

    Article  PubMed  CAS  Google Scholar 

  2. De Cooman SDMN, Dewulf BB, Carette R, Deloof T, Sosnowski M, De Wolf AM, Hendrickx JFA: Desflurane consumption during automated closed-circuit delivery is higher than when a conventional anesthesia machine is used with a simple vaporizer-O2-N2O fresh gas flow sequence. BMC Anesthesiol. 2008, 8: 4-10.1186/1471-2253-8-4.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Hendrickx JF, Van Zundert AA, De Wolf AM: Sevoflurane pharmacokinetics: effect of cardiac output. Br J Anaesth. 1998, 81: 495-501. 10.1093/bja/81.4.495.

    Article  PubMed  CAS  Google Scholar 

  4. Vagts DA, Lockwood GG: The uptake of sevoflurane during anaesthesia. Anaesthesia. 1998, 53: 862-866. 10.1046/j.1365-2044.1998.00487.x.

    Article  PubMed  CAS  Google Scholar 

  5. Peyton PJ, Fortuin M, Robinson GJ, Stuart-Andrews C, Pierce R, Thompson BR: The rate of alveolar-capillary uptake of sevoflurane and nitrous oxide following anaesthetic induction. Anaesthesia. 2008, 63: 358-363. 10.1111/j.1365-2044.2007.05355.x.

    Article  PubMed  CAS  Google Scholar 

  6. Stuart-Andrews C, Peyton P, Humphries C, Robinson G, Lithgow B: Continuous measurement of multiple inert and respiratory gas exchange in an anaesthetic breathing system by continuous indirect calorimetry. J Clin Monit Comput. 2009, 23: 41-49. 10.1007/s10877-009-9161-4.

    Article  PubMed  Google Scholar 

  7. De Cooman S, Lecain A, Sosnowski M, De Wolf AM, Hendrickx JF: Desflurane consumption with the Zeus during automated closed circuit versus low flow anesthesia. Acta Anaesthesiol Belg. 2009, 60: 35-37.

    PubMed  CAS  Google Scholar 

  8. Robinson GJ, Peyton PJ, Terry D, Malekzadeh S, Thompson B: Continuous measurement of gas uptake and elimination in anesthetized patients using an extractable marker gas. J Appl Physiol. 2004, 97: 960-966. 10.1152/japplphysiol.01197.2003.

    Article  PubMed  Google Scholar 

  9. Coppens MJ, Versichelen LF, Mortier EP, Struys MM: Do we need inhaled anaesthetics to blunt arousal, haemodynamic responses to intubation after i.v. induction with propofol, remifentanil, rocuronium?. Br J Anaesth. 2006, 97: 835-841. 10.1093/bja/ael254.

    Article  PubMed  CAS  Google Scholar 

  10. Singaravelu S, Barclay P: Automated control of end-tidal inhalation anaesthetic concentration using the GE Aisys Carestation. Br J Anaesth. 2013, 110: 561-566. 10.1093/bja/aes464.

    Article  PubMed  CAS  Google Scholar 

  11. Body SC, Fanikos J, DePeiro D, Philip JH, Segal BS: Individualized feedback of volatile agent use reduces fresh gas flow rate, but fails to favorably affect agent choice. Anesthesiology. 1999, 90: 1171-1175. 10.1097/00000542-199904000-00033.

    Article  PubMed  CAS  Google Scholar 

  12. Hendrickx JFA MP, De Cooman S, Van Zundert T, De Wolf AM: Vaporizer-fresh gas flow sequences for sevoflurane in O2/air (abstract). 2012, Washington, USA: A192, ASA annual meeting 2012

    Google Scholar 

  13. Hinz JRN, Schwien B, Popov AF, Mohite PN, Radke O, Bartsch A, Quintel M, Zuchner K: Cost analysis of two anaesthetic machines: “Primus®” and “Zeus®”. BMC Res Notes. 2012, 5: 3-10.1186/1756-0500-5-3. doi:10.1186/1756-0500-5-3

    Article  PubMed  PubMed Central  Google Scholar 

  14. Hendrickx JFA, De Cooman S, Cools C, Van de Velde M, De Wolf AM: Do increased canister costs offset decreased sevoflurane costs when fresh gas flows are reduced?. 2011, Chicago, IL, USA: A 1281, ASA annual meeting 2011

    Google Scholar 

  15. Sulbaek Andersen MP, Nielsen OJ, Wallington TJ, Karpichev B, Sander SP: Medical intelligence article: assessing the impact on global climate from general anesthetic gases. Anesth Analg. 2012, 114: 1081-1085. 10.1213/ANE.0b013e31824d6150.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

No further acknowledgements are applicable. Only intradepartmental funding used.

Presented in part at the ESA meeting, June 4th, Barcelona, 2013.

This work has been supported by departmental funds only.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jan FA Hendrickx.

Additional information

Competing interests

The only author who has competing interests is Dr. Hendrickx.

In the past five years, Dr. Hendrickx has received lectures fees, travel reimbursements, and/or research equipment from AbbVie, Baxter, Draeger, GE, Heinen und Lowenstein, Maquet, MEDEC. None of the authors has any other financial or non-financial competing interests.

Authors’ contributions

SDC and JH conceived the idea; SDC collected and processed the data; all authors contributed to data review, analysis, manuscript preparation, and all gave their final approval.

Authors’ original submitted files for images

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver ( https://creativecommons.org/publicdomain/zero/1.0/ ) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

De Cooman, S., Hendrickx, J.F., Peyton, P.J. et al. Agent consumption with the Zeus® in the automated closed circuit anesthesia mode with O2/air mixtures. BMC Res Notes 7, 469 (2014). https://0-doi-org.brum.beds.ac.uk/10.1186/1756-0500-7-469

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/1756-0500-7-469

Keywords