Figure 2From: A validation study: assessing the reliability of the hand held StatStripXPress lactate meter to test lactate in amniotic fluidA Bland- Altman plot showing the agreement between the held meter StatstripXPress and the ABX Pentra 400. The mean difference between hand held meter and lab instrument was -1.1 mmol/L (SD 3.0 mmol/L) across the range 7.7 mmol/L to 20 mmol/L.Back to article page